CSQ Viewer v2.3

Change emissivity, distance, and humidity on FLIR files

It is easy to leave a thermal camera on its default 0.95 emissivity and only notice back at your desk. Because FLIR .csq, .seq, .fff, and radiometric .jpg files store raw sensor readings separately from the scene settings, you can adjust emissivity, object distance, and relative humidity right in your browser.

How to adjust parameters

  1. Open the viewer and upload your radiometric file.
  2. Click Parameters next to the temperature scale dropdown. The panel starts with the values saved by the camera.
  3. Drag the panel by its header if you want to place it beside the image.
  4. Change emissivity (0.01 to 1), object distance (in meters), or relative humidity (0 to 100 %). The preview, color bar, and pixel readout update live.
  5. Use the reset button next to any field to go back to the camera's original value. Closing the panel keeps your adjusted settings active.

Exports use your adjusted values

Once you change a parameter, every export from the viewer picks it up:

  • PNG, batch ZIP, and MP4: rendered from the recalculated temperatures and scale.
  • JSON: exports the recalculated temperature grid in degrees Celsius.
  • FFF and AI-Upscaling: writes your updated emissivity, distance, and humidity into the radiometric header while keeping the raw sensor counts intact.

Your overrides are also stored in the page URL, so reloading the tab or sharing the link keeps the same settings.

Typical emissivity values

Exact values depend on surface finish, angle, and wavelength, so treat these as rough starting points for LWIR cameras (7.5–14 µm):

MaterialEmissivity
Human skin0.98
Water, ice0.95 – 0.98
Matte black paint, black electrical tape0.95
Concrete, brick, plaster, wood0.90 – 0.95
Asphalt, rubber, dark plastics0.90 – 0.95
Glass (measured surface, not through it)0.85 – 0.92
Oxidized steel or iron0.65 – 0.85
Bare or polished metal0.05 – 0.20

Shiny metals have very low emissivity and mostly act like mirrors in the infrared. Even with the right emissivity number, reflected heat from nearby objects or the sky will dominate the reading.

Frequently asked

Can I change emissivity after taking a thermal image?

Yes, if the file is radiometric (FLIR .csq, .seq, .fff, or an unedited radiometric .jpg). Those formats store the raw 16-bit sensor data separately from the scene parameters, so temperatures can be recalculated anytime with different emissivity, distance, or humidity values.

Does changing parameters modify my original file?

No. Your uploaded file stays unchanged. Exports like PNG, MP4, and JSON are rendered with your adjusted settings, and downloading an .fff file saves a new radiometric frame with the updated parameters written into its header.

What happens when I lower or raise emissivity?

Lowering emissivity assumes the surface emits less infrared radiation itself and reflects more of its surroundings. For an object warmer than the background, lowering emissivity increases the calculated temperature; for a colder object, it lowers it.

Why do object distance and humidity matter?

Air between the camera and the target absorbs some infrared radiation, mostly due to water vapor. At a meter or two the effect is small, usually tenths of a degree, but over longer distances in humid conditions it becomes noticeable.

Open a thermal file

Guides

If you need help working with .CSQ files or thermal images and videos in general, feel free to contact me.